3D-printed lung phantom reproduces an embolism for scanner tests

A blood clot lodged in one lobe of the lung is a problem doctors image twice: first with a CT scan to find it, then with nuclear medicine to follow it. A team at the University Hospital of Brest and the University of Western Brittany has now printed a lung that carries the clot on purpose, so that both kinds of scanner can be tested on the same object.
The model started from the lung insert of a commercial thorax phantom, a anatomical test object used to calibrate scanners. The team scanned that insert with CT, mapped anonymized lung data onto it through rigid and deformable registration, and designed a simulated embolism in the superior segment of the left lower lobe, about 5% of lung volume. Two screw plugs let the healthy cavity and the clot compartment be filled separately with contrast agent or radiotracer. Because the model had to hold radioactive solutions, the team chose selective laser sintering to print the nested cavity as a single watertight block. The part, measuring 14.3 x 24.1 x 10.4 cm, was printed in Nylon 12 on a Formlabs Fuse 1+ system at 0.11 mm layers, and it passed a leak test.
Filled with polystyrene beads to mimic lung density and moved by a motorized platform that simulated breathing, the model was imaged with technetium-99m for scintigraphy and gallium-68 for PET. A radiologist and two nuclear medicine physicians judged the images highly realistic across dual-energy CT iodine mapping, scintigraphy and PET. Reading blind, both physicians estimated the blocked share at one segment, or about 5%. That share matters: the pulmonary vascular obstruction index, the part of the lung receiving no blood flow, has been identified as an independent predictor of recurrence in unprovoked embolism.
The authors also report limits. The printed wall is 72% larger by volume than the original insert's, and the clot compartment measures 7.6% of the prototype's volume against the 5% designed. Nylon walls are visible on CT, so the blocked share could not be quantified from CT images. The cost comes to €770, including labor, printing, material and shipping; Formlabs supplied the print free of charge, so the figure reflects estimated market rates.
For radiology and nuclear medicine teams, the immediate gain is a reference object: a lung whose clot size is known in advance, on which a CT scan and a nuclear scan can be compared directly. The authors do not announce a timeline for wider use or a commercial version of the model.