Voltage-switched metalens lets microscope change depth

A tiny blood vessel near the iris of a mouse’s eye comes into focus, and the picture changes without anything in the microscope moving. The switch is made by a voltage, and it takes 0.25 seconds. A team led by Junsuk Rho and Chulhong Kim at POSTECH built the system, and the study appears in Light: Science & Applications.
Photoacoustic microscopy sends laser light into biological tissue. The light is absorbed, which generates ultrasonic waves, and those waves are detected to build an image. Focusing the light tightly gives micrometer-scale resolution and shows fine vascular structures, but it also shortens the depth over which the image stays sharp. Earlier attempts to widen that depth used Bessel beams, which keep a narrow central focus over a longer range, but their side lobes add background signals and reduce contrast.
The POSTECH design avoids both problems by working on the polarization of light. A dual-focus metalens focuses light at two different axial positions depending on whether it is right- or left-circularly polarized. An electrically driven liquid-crystal cell, the same type of liquid crystal used in smartphone displays, changes the polarization of the incident light according to the applied voltage. Alternating between the two focal modes gives an overall axial imaging range of about 1.2 mm, with lateral resolutions of about 3.1 μm and 3.7 μm. A ring-shaped ultrasonic transducer keeps the optical and acoustic axes aligned in a compact setup.
In experiments on mice with chemically induced corneal injury, the shallow mode clearly showed newly formed blood vessels in the cornea. The deep mode selectively resolved vessels near the iris, about 2.2 mm below the surface. These are lab results in animals. The study reports no tests on humans, and the team does not describe a timeline for clinical use.
For eye care, the appeal is practical: clinically relevant vessels lie at very different depths, and diseases such as diabetic retinopathy require accurate assessment across several regions of the eye. If the approach holds up beyond the laboratory, an imaging system that changes focal depth electronically could replace moving optics, which Rho says could make systems “more compact and stable” for both medical and industrial imaging. The next step the team points to is that broader use, which has yet to be tested.